Page 16 · TC-3 Temperature Controller User’s Manual
Menu Description Factory
Default
Cable R The resistance of the cable and
connectors, in ohms. This setting allows
for accurate voltage measurement at
the TEC by removing the voltage loss of
the cable.
0.0080Ω
AutoTune PID AutoTune PID Function
See using the AutoTune Function later
in this manual for details on this item.
Sensor Coeffs Sensor Coefficients Menu
ThermA The A term in the Steinhart-Hart
equation.
1.12924E-03
ThermB The B term in the Steinhart-Hart
equation.
2.34108E-04
ThermC The C term in the Steinhart-Hart
equation.
0.87755E-07
RTD A The A term in the RTD equation.
3.98480E-03
RTD B The B term in the RTD equation.
-0.58700E-06
RTD C The C term in the RTD equation.
4.00000E-12
RTD R
0
The R
0
term (in Ω) in the RTD
equation.
100.00
AD590 M The slope term in the AD590
correction equation.
1.00000
AD590 B The offset term, in °C, in the AD590
correction equation.
0.000
LM335 M The slope term in the LM335
correction equation.
1.00000
LM335 B The offset term, in °C, in the LM335
correction equation.
0.000
Sens Cal M The slope term for user sensor
calibration of RTD or thermistor
measurements.
1.0000
Sens Cal B The offset term for user sensor
calibration of RTD or thermistor
measurements.
0.00 Ω